EM field distribution and SAR in a Human Head with MRI Coil

In  Ultra-High Field MRI  systems, e.g. for B0 beyond 7 T, the performance of the RF coils is essential to provide a great image quality with a reasonable B1-field homogeneity in the imaging area, low SARs in the  biological  tissues  and  a  good  signal  to  noise  ratio (SNR). Despite the many advantages in increasing the B0-field strength (higher resolution and SNR, reduced scan time), operating at higher frequencies adds significant technical complexity to NMR experiments and consequently to the RF coils design. As frequencies increase, the wavelength shortens (60 cm in the vacuum and 8.5 cm inside the head at 500 MHz), and becomes comparable to the electrical dimensions of the head/body and the RF coil. RF fields interact more strongly with human tissues and wave behaviour of the B1-field should affect strongly its homogeneity inside the highly permittive and lossy tissues. While B1 inhomogeneities cause non uniform intensity distribution in MR images, the RF electric field counterpart, potentially inhomogeneous too, should expose biological tissues to an excessive RF power deposition with induced local heating. Coupled to detailed anatomical human models, computational methods offer an indispensible tool for the investigatation of the distribution of the magnetic field or the maximum local SAR values inside the tissues and for the design and evaluation of the RF coils' performance.

This article provides a summary of the work carried out by CEA Saclay, France using CST MICROWAVE STUDIO® (CST MWS) and the HUGO Dataset with the permission and courtesy of Xavier Hanus and his colleagues[1].

 

 


CST MWS Setup with HUGO Head model and the definition of Lines for Field Evaluation
Figure 1: CST MWS Setup with HUGO Head model and the definition of Lines for Field Evaluation

CST MWS was used in conjunction with the 3D anatomical data set HUGO to simulate the time varying fields of an RF coil loaded by a human head section of the HUGO model with a mass of 7.81kg. A Dell 650 station was used with CPU P4 Xeon 2.8 GHz, 2 Go of RAM and SCSI HD. The model setup is shown in Figure 1 with the lines defined for the fields' extraction. 


Details of the meshing of the anatomical 3D human model inserted in the coil in axial (a,b) and sagital (c,d) planes
Figure 2: Details of the meshing of the anatomical 3D human model inserted in the coil in axial (a,b) and sagital (c,d) planes

For the example presented here, the total number of mesh cells over the entire volume of calculation is 1,982,178 as shown in Figure 2, with about 1,100,000 cells for the tissues corresponding to a 3 by 3 by 5 mm average resolution (roughly 18 voxels per 1 cm3 of tissue).


Locations of the 4-ports (a) and detail of the meshing in the area of port 1 (b,c)
Figure 3: Locations of the 4-ports (a) and detail of the meshing in the area of port 1 (b,c)

All the outer surfaces of the calculation volume are set to open boundaries with PML layers. As shown in Figure 3, four discrete ports have been defined at the bottom of the coil at 90° in the axial plane, with adjustable impedance and lumped capacitance in parallel for better matching when loading with the head.


S-Parameters for the head loaded Coil
Figure 4: S-Parameters for the head loaded Coil

For the high meshing intensive calculations, the excitations are performed over 0-800 MHz through the ports 1 and 3 in quadrature. As a result of its polarization, this mode, called B1, is not transmitted between the ports in quadrature as indicated by the transmission S 3,1 for the loaded coil at 480 MHz as shown in Figure 4.


Post-combination of the  1-port linear excitations Lin1P1 and Lin1P3 into the 1 watt circular H-field
Figure 5: Post-combination of the 1-port linear excitations Lin1P1 and Lin1P3 into the 1 watt circular H-field


The  fields  and  power  loss  density  are  obtained  at  a  specific frequency  for  each port excitation.  As shown Figure 5, the B1-mode presents a magnetic vector oriented linearly along -Oy when excited (Lin1P1) by the port 1 along x (cf. Figure 3.a and 5.a). The  results  are  then combined  as  a  post-processing  step  with  specific  amplitudes  compensating  for reflections (|S11| and |S33|), and phase-shifting (90°) in order to create the circularly polarized excitation. When animated with wt, the magnetic vector field appears rigorously circular in the axial plane for the empty coil (Fig. 5.a3) but quite disturbed in the head (Fig. 5.b3). In MRI, only the strictly circular component of the rotating magnetic field is effective on the spins excitation and participates to the formation of the images. 


Magnitude and phase of H+1 along x and y for empty and loaded coils
Figure 6: Magnitude and phase of H+1 along x and y for empty and loaded coils

Figure 6 shows the spatial variation of the magnitude (left) and phase (right) of the H+1 vector along the transverse axes for both empty and loaded coil.  |H+1| is enhanced with a sharp focusing effect at the centre of the brain : up to 1.2 A/m for the loaded coil compared to the 0.95 A/m uniform profile of the unloaded coil (Fig. 6.a). When loading with the head, phases are not symmetrically distributed over -45° and a strong phase-shifting is acting at the high/low permittivity interfaces of the tissues white and grey matter/bones and fat (positions 150/250 mm curve blue for the x-direction Fig. 6.b).


1g local SAR plots on xy and yz planes
Figure 7: 1g local SAR plots on xy and yz planes

Numerical investigations of SAR and B1 profiles provide crucial information about head loaded coils designed for UHF-MRI. |B1| and SAR values normalized to an effective power injected of 1 Watt CW give the necessary amplification for given tip angles  and  allow  the  calculations  of  expected  SARs  for  MRI  typical  pulse  sequences  in  accordance  to  guidelines.   More studies have to be carried out in order to improve the performances of the coil. Such studies include the increasing of the Q (Qloaded=40), the influence of the ports disposition, the investigation of multi-excitation scheme, composing a better uniform and circular rotating B1+ becoming a major challenge in UHF-MRI with the inherent reduction of the wavelength.

[1] X. Hanus, M. Luong, F. Lethimonnier, "Electromagnetics Fields and SAR Computations in a Human Head with a Multi-port Driven RF Coil at 11.7 Tesla", Proc. Intl. Soc. Mag. Reson. Med. 13 (2005).


CST Article "EM field distribution and SAR in a Human Head with MRI Coil"
last modified 21. May 2008 12:01
printed 7. Oct 2008 3:25, Article ID 259
URL:

All rights reserved.
Without prior written permission of CST, no part of this publication may be reproduced by any method, be stored or transferred into an electronic data processing system, neither mechanical or by any other method.

Other Articles

Radar Cross Section and Surface Current Simulation for a Helicopter

Radar Cross Section and Surface Current Simulation for a Helicopter
We present the RCS and surface current simulation for an electrically large helicopter. The simulation is done using a plane wave excitation at 7GHz. The helicopter is approximately 180 wavelengths in size. The calculation is performed with the Integral Equation solver of CST MICROWAVE STUDIO® using the Multilevel Fast Multipole Method. The solver is very accurate and efficient in terms of complexity for simulating electrically large structures. Read full article..

Electrical Fast Transient/Burst (EFT/B) Susceptibility of an LV Circuit Breaker

Electrical Fast Transient/Burst (EFT/B) Susceptibility of an LV Circuit Breaker
The coupled voltage at component level has been simulated using CST MWS for two PCB releases of the same Low Voltage Circuit Breaker (LV-CB) electronic control scheme. The results show a good agreement with actual measurements. Read full article..

Simulation of 650 GHz Backward Wave Oscillators

Simulation of 650 GHz Backward Wave Oscillators
In this acticle the simulation of novel interdigital backward wave oscillators with CST PARTICLE STUDIO® is described. The mutual coupling of the charged particle movement and transient electromagnetic fields is taken into account by a sophisticated Particle in Cell algorithm. The resulting output power is in excellent agreement with theoretical values. Results are presented with the courtesy and permission of Teraphysics Corporation, USA. Read full article..

Phase Center Computation of a Corrugated Horn

Phase Center Computation of a Corrugated Horn
For a customer given corrugated horn the phase center has been computed using CST MICROWAVE STUDIO® and compared with measurement data. The simulated results are in very good agreement with measurements. Measurements were provided with courtesy and permission of Kathrein Werk KG, Rosenheim, Germany. Read full article..

RFID Transponder operating at 13.56 MHz

RFID Transponder operating at 13.56 MHz
Radio Frequency Identification Systems (RF-ID) are widely used and are thus one of the fastest growing sectors of todays radio industry, allowing advanced solutions for a variety of applications in the area of authentication, ticketing, access control, supply management, etc. One of the most common band allocated to RFID systems is 13.56 MHz. For this application example operating at this particular frequency band we have chosen a transponder inlay which was created using the ACIS based solid modeler of CST MICROWAVE STUDIO® The frequency domain solver of CST MICROWAVE STUDIO® has been applied to accurately predict the input impedance, followed by a lumped element based equivelent circuit derivation to describe the impedance versus frequency. Read full article..

Back