Modeling Temperature Effects of RF Thermoablation in a Human Liver using the bioheat formulation in CST DE
Approximately 70% of people who contract cancer of the liver are unable to undergo surgical treatment and are considered to have inoperable cancer. Among the most successful minimally invasive treatments is the direct application of energy using treatments such as Radio Frequency Thermoablation (RFT).
The goal of RF thermoablation is to deposit RF energy in a tumour. RF electrodes are inserted into the tumour and under the guidance of either ultrasound, CT or MRI, RF energy is deposited into the tumour. The subsequent heating of the region destroys the tumour and tissue in a small circumferential region around the tip of the electrode used for depositing the RF energy. The amount of deposited energy is carefully controlled in order to minimize damage to healthy tissue.
Figure 1: Simulation setup: Catheter is inserted into the abdomen of the HUGO anatomical model
Determining the temperature distribution in the liver by means of EM simulation. The finite integration technique is very well suited to simulating the EM interaction between thin metallic radiators and heterogeneous voxel based anatomical models. In order to determine the temperature distribution in the liver due to the electromagnetic fields in the body, a co-simulation was performed using CST MWS and CST EMS. The simulation set-up consists of an electrode, a voxel model of the human body and a reference (grounding) plate.
This set-up is demonstrated in Figure 1, where a cut view of the model shows the organs and the location of the inserted electrode. The electrode is modeled according to the dimensions of the Elektrotom Hitt 106 device (www.integra-ls.com), and consists of a perfect electric conductor (PEC) and lossless Teflon. Two different body models are used for comparative analysis; these are the HUGO body model, based on anatomical data of the Visible Human Project®, and also a heterogeneous anatomical model (UAq ALES) developed at UAq EMC Laboratory, University of L'Aquila, Italy (http://orlandi.ing.univaq.it/UAq_Laboratory/index.html).
Figure 2: Modeling details of the Catheter
Since the heating is localized, only a section of the torso was simulated. Some of the details of the catheter (electrode) are shown in Figure 2. A 40 W, Gaussian signal with center frequency 375 MHz was used as excitation.
Figure 3: Electric field inside the HUGO model
An animation of the Electric field inside the HUGO model is shown in Figure 3. Due to the dielectric and conductive properties of the human tissue, the fields are concentrated around the catheter and the skin touching the groundplane. In Figure 4, the power loss density inside the HUGO model (a) and the UAq model (b) is shown.
The slight differences in the distribution of the power loss density are considered negligible, since the two models are different from each other and it is impossible to have identical cross sections for both models.
Figure 4: Power Loss Density calculated in (a) HUGO body model and (b) U Aq body model
Using co-simulation with CST EM STUDIO™ (CST EMS), it is possible to calculate the heating in the tissue, due to an induced current, which includes cooling effects from blood flow as described by the bioheat equation. In Figure 5, a cross section of the temperature distribution in the liver of the HUGO model is shown.
Figure 5: Temperature distribution inside the HUGO model
In this study, a co-simulation using CST MWS and CST EMS was performed in order to determine the localized heating inside a tumor when exposed to an RF signal.
Due to the heterogeneous, voxel nature of the model, a time domain solver was used. In the solution of heterogenous body models, the time domain solution requires significantly less memory, than a frequency domain method. In order to verify results, the simulation was performed on two different body models. The results were found to be in good agreement.
CST Article "Modeling Temperature Effects of RF Thermoablation in a Human Liver using the bioheat formulation in CST DE"
last modified 8. Jan 2009 1:23
printed 10. Feb 2012 10:56, Article ID 387
URL:
All rights reserved.
Without prior written permission of CST, no part of this publication may be
reproduced by any method, be stored or transferred into an electronic data processing system, neither mechanical or by any other method.
Article ID: 387
Last modified: 8. Jan 2009 1:23
Other Articles
This article shows a common approach to the simulation of electromechanical problems. A set of 3D electromagnetic field simulations using CST EM STUDIO® (CST EMS) are carried out on a linear actuator in order to obtain its electromagnetic characteristics for use in Simulink® where both the electrical and mechanical behaviour can be simulated.
Read full article..
Christof Thalhammer, Berlin Ultrahigh Field Facility, Max-Delbrueck-Center
At the Berlin Ultrahigh Field Facility we explore the advantages and needs of Magnetic Resonance Imaging (MRI) at 7 Tesla. Apart from the magnet and the gradient system, the radio frequency (RF) coils form an essential part of the MR system and are topic of intense research. Ultrahigh field systems are not yet in clinical use and the number of commercial coils for these systems is very limited, hence the coil design is an important part of our facility’s work. Since higher magnetic fields require higher frequencies, one has to deal with stronger interaction of the RF fields with the human tissue. Therefore conventional coil designs established at lower magnetic field strengths cannot be directly applied at 7.0 T which asks for new concepts and developments. To accelerate and streamline the design of new coils, our group recently started to use CST Studio Suite 2010. Together with detailed voxel models of human bodies (the “Virtual Family” by the ITIS Foundation), we use the CST package to simulate the distribution of the magnetic field and the RF power distribution in the patient’s body with the ultimate goal to optimize our coil designs for reasons of B1-homogeneity and local SAR distribution. Some of the first results will be shown to illustrate the application of CST Studio Suite in one of our projects, i.e. the design of a multi channel transmit/receive coil array for cardiovascular MRI at 7.0 Tesla.
Read full article..
Radio Frequency Identification Systems (RFID) are widely used and allow advanced solutions for a variety of applications in the area of authentication, ticketing, access control, supply management, parking, payment, vending,etc. The example presented here is a RFID Readercoil "P81" from Legic Ident Systems and was modeled and solved using the frequency domain solver of CST MICROWAVE STUDIO® (CST MWS). The sensitivity of the computed complex input impedance with respect to substrate tolerances is computed and was compared to measurement data.
Read full article..
The telecommunications sector is making great advances aimed at delivering an even stream of high tech devices, covering the significant consumer demands in this sector. EM simulation is increasingly becoming an indispensable tool in the design flow, not only on the antenna level but also on the phone and environmental levels. This article compares simulated results with measurements for several steps in the phone design chain.
Read full article..
A portion of a backplane, designed with the Cadence Allegro® layout tool, is imported into CST MICROWAVE STUDIO®. This section consists of a differential pair with vias which go through the board. The structure was analyzed in Microwave Studio. The simulation results demonstrate the impact of backdrilling the signal vias to improve the signal integrity performance. A detailed SPICE model of the transmission path is created and its accuracy is verified.
Read full article..